ISO 24427:2025
ISO 24427:2025 Radiological protection - Medical proton accelerators - Requirements and recommendations for shielding design and evaluation
standard byInternational Organization for Standardization , 09/01/2025
standard byInternational Organization for Standardization , 09/01/2025
This document is applicable to the radiation shielding design and evaluation work for medical proton accelerators of proton energies ranging from 70MeV to 250MeV, with subsystems such as beam transport system and nozzle components.
u000bThe radiation protection recommendations given in this document cover the aspects relating to regulations, shielding design goals and other design criteria, role of the manufacturers, of the radiation protection officer or qualified expert, the medical physicist, the licensee and interactions between them, sources and radiations around a proton accelerator, shielding for accelerators and its subsystems (including shielding materials and transmission values, calculations for various room configurations, duct impact on radiation protection) and the radiological measurements.
u000bFLASH proton therapy is not covered by this document.
u000bNOTE1 AnnexA provides a list of the most used Monte-Carlo codes for shielding calculation.
u000bNOTE2 AnnexB provides the analytical methods and the corresponding necessary data for shielding calculation.
u000bNOTE3 AnnexC provides a set of examples on shielding calculation of barriers, maze and skyshine problems.
u000bNOTE4 AnnexD provides radiation shielding consideration on special topics.